氣相色譜48個常見問題及注意事項(下)

發布日期:2021-09-07
核心提示:廿五、色譜柱的常用填充方法有哪些?答:固定相填充的好壞,將直接影響柱效率。通常多用泵抽填充法,即把色譜柱的一端塞上玻璃棉
答:gudingxiangtianchongdehaohuai,jiangzhijieyingxiangzhuxiaolv。tongchangduoyongbengchoutianchongfa,jibasepuzhudeyiduansaishangbolimian,jiezhenkongbeng,lingyiduanjieyiloudou,zaichouxixiajiarugudingxiang,bianzhuangbianqiaodasepuzhu,zhigudingxiangbuzaijinruweizhi。zhuanghaohou,saishangbolimian。zhuangzhuyaoqiuyaotianchongdejunyun,jinmi,qiejiyoukongxi。廿六、新裝填的色譜柱為什麼要老化一段時間才能使用?答:裝zhuang填tian好hao的de色se譜pu柱zhu,連lian接jie於yu儀yi器qi上shang後hou,應ying先xian試shi壓ya,試shi漏lou,而er後hou在zai恒heng定ding的de溫wen度du下xia用yong載zai氣qi吹chui洗xi數shu小xiao時shi後hou承cheng受shou分fen析xi,一yi般ban稱cheng此ci為wei柱zhu子zi的de老lao化hua過guo程cheng。老lao化hua的de目mu的de是shi把ba固gu定ding相xiang的de殘can存cun溶rong劑ji,低di沸fei點dian雜za質zhi,低di分fen子zi量liang固gu定ding液ye等deng趕gan走zou,使shi記ji錄lu器qi基ji線xian平ping直zhi,並bing在zai老lao化hua溫wen度du下xia使shi固gu定ding液ye在zai擔dan體ti表biao麵mian有you一yi個ge再zai分fen布bu過guo程cheng,從cong而er塗tu得de更geng加jia均jun勻yun牢lao固gu。裝zhuang填tian好hao的de色se譜pu柱zhu,經jing過guo老lao化hua一yi段duan時shi間jian後hou,柱zhu效xiao及ji性xing能neng均jun穩wen定ding了le,這zhe樣yang才cai可ke使shi用yong。答:色譜柱失效主要表現為色譜分離不好和組分保留時間顯著變短。色譜柱失效的主要原因是:對氣固色譜來說是固定相的活性或吸附性能降低了,對氣液色譜來說,是使用過程中固定液逐漸流失所致。老化的目的:氣相色譜柱的固定相通常是以塗覆的形式分布在柱管管壁內側(毛細管柱)或載體表麵(填充柱)shangde,duiyuyigenxindeqixiangsepuzhu,waicenggudingxiangyuzaitidejiehewangwangjiaoruo,zaigaowenxiashiyonghuihuanmanliushi,zaochengjixianqifuhezaoshengshenggao,weilebimianzheyixianxiangfasheng,keyiyuxianzaijiaogaowenduxia(一般為色譜柱的耐受溫度)加熱一段時間,使結合較弱的固定相揮發出去,從而使後麵的分析不受幹擾。此外,對使用時間較長的氣相色譜柱可進行老化操作,可以除去色譜柱中殘留的汙染物。將色譜柱柱溫升至一恒定溫度,通常為其溫度上限。特殊情況下,可加熱至高於操作溫度10-20 ºC 左右,但是一定不能超過色譜柱的溫度上限,那樣極易損壞色譜柱,此外不要將程序升溫的速度設定的太慢。當達到老化溫度後,記錄並觀察基線。比例放大基線,以便容易觀察。初始階段,基線應持續上升,在到達老化溫度後5-10 分鍾開始下降,並且會持續30-90 分鍾。當達到一個固定的值後,基線就會穩定下來。如果在2-3 小時後基線仍無法穩定或在15-20 分鍾後仍無明顯的下降趨勢,那麼有可能係統裝置有泄漏或汙染。遇到這樣的情況,應立即將柱溫降至40ºC 以下,盡快地檢查係統並解決相相關的問題。如果還是繼續地老化,不僅對色譜柱有損害,而且始終得不到正常穩定的基線。另外,老化的時間也不宜過長,不然會降低色譜柱的使用壽命。一般來說,塗有極性固定相和較厚塗層的色譜柱老化時間較長,而弱極性固定相和較薄塗層的色譜柱所需時間較短。而色譜柱的老化方法各不相同,具體步驟請參閱隨柱子的操作說明書。如果在色譜柱沒有與檢測器連接就進行老化,那麼老化後,譜柱末端部分可能已被破壞。要先把柱末端10-20cm 部分截去,再將色譜柱連接到檢測器上。溫度限定是指色譜柱能夠正常使用的應用溫度範圍。如果操作溫度低於色譜柱的溫度下限,那麼分離效果和峰形都不會很理想。但這樣對色譜柱本身並無什麼損害。溫度上限通常有兩個數值。數值較低的是恒溫極限。在此溫度下,色譜柱可以正常使用,而且無具體的持續時間限製。較高的數值是程序升溫的升溫極限。該溫度的持續時間通常不多於十分鍾。高於溫度上限的操作則會降低色譜柱的使用壽命。在GC 中使用程序升溫時常常會出現基線漂移的現象,這種現象通常有以下幾個原因:色譜柱流失、進樣墊流失、進樣器汙染或檢測器汙染、氣(qi)體(ti)流(liu)速(su)的(de)變(bian)化(hua)。如(ru)果(guo)使(shi)用(yong)高(gao)靈(ling)敏(min)度(du)檢(jian)測(ce)器(qi),即(ji)便(bian)是(shi)微(wei)弱(ruo)的(de)柱(zhu)流(liu)失(shi)或(huo)係(xi)統(tong)汙(wu)染(ran)都(dou)可(ke)能(neng)帶(dai)來(lai)顯(xian)著(zhu)的(de)基(ji)線(xian)漂(piao)移(yi)現(xian)象(xiang)。為(wei)了(le)提(ti)高(gao)定(ding)性(xing)和(he)定(ding)量(liang)分(fen)析(xi)的(de)可(ke)靠(kao)性(xing),應(ying)盡(jin)可(ke)能(neng)的(de)降(jiang)低(di)或(huo)消(xiao)除(chu)基(ji)線(xian)漂(piao)移(yi)。首先把柱子從色譜儀上取下,堵住檢測器的入口,再觀察在程序升溫時基線的漂移情況。如果基線不穩,那麼汙染來自檢測器(解決辦法請參考“如何降低檢測器的汙染”) ;如ru果guo基ji線xian是shi穩wen定ding的de,證zheng明ming檢jian測ce器qi良liang好hao,此ci時shi用yong一yi小xiao段duan熔rong融rong石shi英ying管guan把ba進jin樣yang器qi和he檢jian測ce器qi連lian接jie起qi來lai,走zou一yi個ge升sheng溫wen程cheng序xu,觀guan察cha基ji線xian漂piao移yi情qing況kuang,此ci時shi反fan映ying的de是shi進jin樣yang器qi的de汙wu染ran情qing況kuang,如ru果guo基ji線xian不bu穩wen,可ke以yi確que定ding問wen題ti來lai自zi進jin樣yang口kou(解決辦法請參考“如何降低進樣器的汙染” );如果基線穩定,證明檢測器和進樣口均未被汙染,此時把柱子重新裝上,走同樣的升溫程序,來確定是不是柱子流失帶來的基線漂移。色(se)譜(pu)柱(zhu)上(shang)如(ru)果(guo)有(you)高(gao)分(fen)子(zi)不(bu)揮(hui)發(fa)性(xing)物(wu)質(zhi)殘(can)留(liu),那(na)麼(me)在(zai)程(cheng)序(xu)升(sheng)溫(wen)時(shi)就(jiu)容(rong)易(yi)產(chan)生(sheng)基(ji)線(xian)漂(piao)移(yi),因(yin)為(wei)這(zhe)些(xie)物(wu)質(zhi)的(de)保(bao)留(liu)較(jiao)強(qiang),在(zai)柱(zhu)中(zhong)移(yi)動(dong)緩(huan)慢(man),可(ke)以(yi)采(cai)用(yong)重(zhong)新(xin)老(lao)化(hua)的(de)方(fang)法(fa)將(jiang)這(zhe)種(zhong)強(qiang)保(bao)留(liu)組(zu)分(fen)從(cong)柱(zhu)子(zi)上(shang)趕(gan)出(chu),但(dan)這(zhe)種(zhong)方(fang)法(fa)增(zeng)加(jia)了(le)固(gu)定(ding)液(ye)氧(yang)化(hua)的(de)可(ke)能(neng)性(xing);此外,還可以使用溶劑衝洗色譜柱(衝洗之前請閱讀柱子的使用注意事項,以便選出合適的溶劑);也可以安裝保護柱,這樣可以預防問題發生。如果是進樣器被汙染造成基線漂移,可以通過更換進樣墊、chenguanhemifengquanlaijiejue,tongshiyongrongjichongxijinyangkou,weihuwanbizhihou,yongyiduanrongrongshiyingguanjiangjinyangqihejianceqilianjieqilai,jinyizhenkongyang,yiquerenjinyangqiyijingganjing。 由(you)檢(jian)測(ce)器(qi)帶(dai)來(lai)的(de)基(ji)線(xian)漂(piao)移(yi)通(tong)常(chang)是(shi)由(you)補(bu)償(chang)氣(qi)或(huo)者(zhe)燃(ran)氣(qi)當(dang)中(zhong)少(shao)量(liang)的(de)烴(ting)類(lei)物(wu)質(zhi)引(yin)起(qi)的(de),使(shi)用(yong)高(gao)純(chun)氣(qi)體(ti)淨(jing)化(hua)器(qi)處(chu)理(li)補(bu)償(chang)氣(qi)或(huo)者(zhe)燃(ran)氣(qi)可(ke)以(yi)減(jian)少(shao)這(zhe)種(zhong)基(ji)線(xian)漂(piao)移(yi);使用高純氣體發生器可以改善FID 的基線穩定性;正確的檢測器維護,包括定期的清洗,都可以減少這種漂移。 在使用新柱之前,按照以下方法老化可以使柱流失降到:用高於實驗操作溫度20 或者用色譜柱的操作溫度(使用兩者中較低者)來(lai)老(lao)化(hua),長(chang)時(shi)間(jian)低(di)溫(wen)老(lao)化(hua)相(xiang)對(dui)於(yu)短(duan)時(shi)間(jian)高(gao)溫(wen)老(lao)化(hua)有(you)利(li)於(yu)降(jiang)低(di)色(se)譜(pu)柱(zhu)流(liu)失(shi)。如(ru)果(guo)在(zai)載(zai)氣(qi)當(dang)中(zhong)含(han)有(you)少(shao)量(liang)的(de)氧(yang)氣(qi)或(huo)者(zhe)水(shui)分(fen)或(huo)者(zhe)氣(qi)體(ti)管(guan)路(lu)漏(lou)氣(qi),在(zai)高(gao)溫(wen)條(tiao)件(jian)下(xia),固(gu)定(ding)液(ye)就(jiu)容(rong)易(yi)被(bei)氧(yang)化(hua),從(cong)而(er)造(zao)成(cheng)柱(zhu)流(liu)失(shi),帶(dai)來(lai)基(ji)線(xian)漂(piao)移(yi)。一(yi)旦(dan)固(gu)定(ding)液(ye)被(bei)氧(yang)化(hua),必(bi)須(xu)使(shi)用(yong)高(gao)純(chun)載(zai)氣(qi)老(lao)化(hua)數(shu)小(xiao)時(shi),才(cai)有(you)可(ke)能(neng)使(shi)基(ji)線(xian)趨(qu)於(yu)水(shui)平(ping),這(zhe)種(zhong)對(dui)固(gu)定(ding)液(ye)的(de)破(po)壞(huai)是(shi)無(wu)法(fa)彌(mi)補(bu)的(de),所(suo)以(yi)如(ru)果(guo)有(you)氧(yang)氣(qi)連(lian)續(xu)通(tong)過(guo)色(se)譜(pu)柱(zhu),即(ji)便(bian)進(jin)行(xing)老(lao)化(hua)基(ji)線(xian)也(ye)無(wu)法(fa)降(jiang)到(dao)水(shui)平(ping)。因(yin)此(ci),在(zai)實(shi)驗(yan)過(guo)程(cheng)中(zhong),應(ying)在(zai)氣(qi)體(ti)管(guan)路(lu)當(dang)中(zhong)使(shi)用(yong)高(gao)質(zhi)量(liang)的(de)氧(yang)氣(qi)/水分過濾器,同時用高質量的電子檢漏儀嚴格檢漏。6 NPD溫度過高(使用或環境溫度),氣體不純 ,更換銣珠:避免高溫使用。1 峰伸舌多為色譜柱過載,減小進樣量,使用大容量柱子。在連續進樣的條件下,峰麵積忽大忽小,測定精度不高,原因如下:4 色譜柱,襯管被汙染,清洗襯管,用溶劑(優級純甲醇)清洗色譜柱:更換之(如有必要)。1 襯管,色譜柱被汙染或者襯管,色譜柱安裝不當,存在死體積,注射甲烷,峰若拖尾,則重新安裝。5 樣品流通路線中有冷井,消除路線中的過低溫度區。6襯管或色譜柱中有堆積切割碎屑 清洗更換襯管,切除柱頭10cm6 分流進樣時,初始OVEN過高 降低初始柱溫,使用高沸點溶劑。7吹掃時間過長(不分流進樣) 定義短時間的吹掃程序。1 新安裝的柱子,基線連續向漂移幾分鍾 繼續老化。3 檢測器或GC係統中其他部分有沉積物被烤出來,清洗之。2 使用ECD,TCD氣體泄露引發基線噪音 檢查,維修氣路。3 FID ,NPD ,FPD燃氣流速或燃氣選擇不當 高純燃氣,調整流速。4進樣口被汙染 清洗進樣口,更換擱墊,更換襯管中的玻璃纖維。5毛細管色譜柱被汙染 切除首端10cm,用溶劑清洗色譜柱,更換之。7 減少係統的死體積,比如色譜柱連接要插到位,不分流進樣要選擇不分流結構汽化室。綜上所述要根據具體情況在實驗中摸索,比如降低載氣流速、降低色譜柱溫度又會使色譜峰變寬,因此要看色譜峰型來改變條件。最終目的是達到分離好,出峰時間快。FIDjianceqizuishiheyongyujiancesepuzhulaohuashidejixian。zaishengwenchengxudemoduan,jixianjiangshenggao,ranhoujixianxiajiangzhujianpingwen,cishikeyirenweisepuzhulaohuawancheng。 當色譜柱處於高溫時,柱壽命急劇下降。如果色譜柱老化時超過2小時還有大量柱流失,則將色譜柱冷卻至室溫,辨認柱流失來源如:氧氣滲入、隔墊漏氣和儀器本身的殘留物。柱流失:在色譜柱老化之後做柱流失實驗,不進樣跑一次程序升溫,從50℃開始升溫 10℃/min到色譜柱最高使用溫度,並在最高溫度保持10min chulaideseputujiweizhuliushitu,nazhezhangtugenjinhoukongbaiduibi。ruguozaikongbaiyunxingzhongchanshenglehenduofeng,zesepuzhuxingnenggaibian,zhekenengshiyouyuzaiqizhonghanyouyangqi,yekenengshiyouyuyangpincanliu。ruguoyou GC-MS,則低極性色譜柱的典型流失離子(例如 DB/HP-1 或 5)質/荷比 m/z 將為 207、73、281、355 等,大多數為環矽氧烷。一(yi)般(ban)認(ren)為(wei)柱(zhu)流(liu)失(shi)能(neng)引(yin)起(qi)噪(zao)聲(sheng)和(he)不(bu)穩(wen)定(ding)的(de)基(ji)線(xian)。真(zhen)正(zheng)的(de)柱(zhu)流(liu)失(shi)常(chang)常(chang)有(you)如(ru)同(tong)噪(zao)聲(sheng)狀(zhuang)的(de)正(zheng)向(xiang)漂(piao)移(yi)。看(kan)看(kan)基(ji)線(xian)是(shi)否(fou)向(xiang)上(shang)較(jiao)大(da)漂(piao)移(yi),空(kong)白(bai)有(you)無(wu)峰(feng)流(liu)出(chu)等(deng)。四十六、氫火焰離子化檢測器(FID)火焰熄滅或點不著火的原因分析①冷凝。由於FID燃燒過程中導致水的形成,所以檢測器溫度必須保持高於1 0 0℃,以免冷凝。長時間不開機時,需長時間進行烘烤後再點火。②柱流速過高。若必須使用大內徑柱,可關小載氣流速足夠長時間以使FID點火。③檢查安裝的噴嘴類型是否適合使用的色譜柱,檢查噴嘴是否堵塞。1、氣體鋼瓶是貯存壓縮氣體的高壓容器,其容積一般為40~60 L,最高工作壓力為15MPa(150atm),最低的也在0.6MPa(6atm)以上,標準高壓氣體鋼瓶是按國家標準製造而成,在鋼瓶肩部應有下述標記,即:製造廠、製造日期、氣瓶型號及編號、氣瓶重量、氣體容積、工作壓力、水壓試驗壓力、水壓試驗日期及下次送檢日期等。由於氣體鋼瓶壓力很高,有的氣體有毒或易燃易爆,為了確保安全,避免各種鋼瓶相互混淆,應按規定在鋼瓶外麵塗上特定的顏色,寫明瓶內氣體的名稱。各種氣體鋼瓶的標誌見表1-3。
(1)鋼瓶必須定期送有關部門檢驗,檢驗合格的才能充氣。充一般氣體的鋼瓶3年內必須送檢一次,充腐蝕性氣體的鋼瓶每兩年送檢一次。(2)搬運鋼瓶時,要戴好鋼瓶帽和上、下兩個橡皮腰圈,輕拿輕放,不可在地上滾動、撞擊、摔倒或激烈振動,以防發生爆炸。放置和使用鋼瓶時,必須用架子或鐵絲固定住。 (3)鋼瓶應存放在陰涼、幹燥、遠離熱源的地方,通風良好,避免明火和陽光暴曬。鋼瓶受熱後,氣體膨脹,瓶內壓力增大,易造成漏氣,甚至爆炸。可燃性氣體鋼瓶與氧氣鋼瓶必須分室存放。氫氣鋼瓶最好放置在大樓外的專用小間,以確保安全。(4)使用鋼瓶,除二氧化碳、氨氣外,一般要用減壓閥。各種減壓閥中,除氮氣和氧氣的減壓閥可相互通用外,其他的隻能用於規定的氣體,以防爆炸。安裝減壓閥必須仔細旋妥,通常旋進7圈螺紋(俗稱吃7牙)。易發生聚合反應的氣體(如乙炔、乙烯)必須規定儲存期限,避免久貯。(5)可燃性氣體,如氫氣、乙炔等,鋼瓶的閥門是“反扣”(左旋)螺紋,即逆時針方向擰緊;非燃性或助燃性氣體,如氧氣、氮氣等,鋼瓶的閥門是“正扣’,(右旋)螺紋,即順時針擰緊。(6)絕不可將油、脂或其他易燃物、有機物沾在氧氣鋼瓶上,特別是閥門嘴及減壓閥處,也不得用棉、麻等物堵漏,以防燃燒引起事故。 (7)要yao注zhu意yi保bao護hu好hao鋼gang瓶ping閥fa門men。開kai關guan閥fa門men時shi,首shou先xian弄nong清qing方fang向xiang,再zai緩huan慢man旋xuan轉zhuan,否fou則ze會hui使shi螺luo紋wen受shou損sun。開kai啟qi閥fa門men時shi,人ren應ying站zhan在zai減jian壓ya閥fa的de另ling一yi側ce,以yi防fang減jian壓ya閥fa衝chong出chu被bei擊ji傷shang,每mei次ci用yong後hou應ying完wan全quan關guan閉bi閥fa門men。(8)貯存可燃性氣體的鋼瓶要有防回火裝置。有的減壓閥已有此裝置;也可在氣體導管中填裝細鐵絲網防止回火;在導氣管路中加接液封裝置也可有效地起到保護作用。(9)不得將鋼瓶內的氣體全部用完,一定要保留0.05 MPa以上的殘餘壓力(減壓閥表壓)。可燃性氣體(如乙炔)應剩餘0.2~0.3 MPa,氫氣應保留2 MPa,以防重新充氣時發生危險。鋼瓶要隨用隨關,勤檢查。(10)一旦發生閥門漏氣,應立即將鋼瓶移至室外,以防在室內發生事故。(1)氫氣。若從鋼瓶中急劇地放出,即使沒有火源存在,有時亦會著火。氫氣和空氣混合物的爆炸範圍很寬,當含氫氣4.0%~75.6%(體積比)時,遇火即會爆炸。因此,氫氣要在通氣良好的地方使用,或用排氣筒盡量地把室內氣體排到室外。(2)乙炔。極易燃,且燃燒溫度很高,有時還會發生分解爆炸。乙炔與空氣混合時的爆炸範圍為含乙炔2.5%~80.5%(體積比)。因此,要嚴禁煙火,防止漏氣。隻zhi要yao色se譜pu係xi統tong受shou到dao高gao沸fei點dian物wu質zhi的de汙wu染ran,特te別bie是shi在zai進jin樣yang口kou,就jiu可ke以yi預yu料liao色se譜pu性xing能neng會hui變bian差cha。分fen析xi人ren員yuan應ying當dang進jin行xing儀yi器qi的de曰yue常chang維wei護hu,包bao括kuo定ding期qi更geng換huan進jin樣yang隔ge墊dian、清洗和老化進樣口內襯管等,必要時可將接於進樣口一端的毛細管色譜柱截去0.5-1m。ruguoyijiuchuxiansepuxingnengjiangdiheguifengwenti,kenengxuyaoqingxijinyangkoudejinshubiaomian。maoxiguansepuzhushikekaobingyiyushiyongde,danshi,weilebaozhenglianghaodefenlixingneng,xuyaozhuyixialietedingcaozuo: (1)毛細管柱和色譜爐壁之間的接觸可以影響色譜性能和色譜柱壽命; (4)再次加熱色譜爐之前,應當先用載氣衝洗色譜柱15 rnin; (5)應當使用脫氧管除去載氣中的痕量氧氣,脫氧管應當定期更換; (6)無論色譜爐是否在加熱,都需要有載氣流經色譜柱。
編輯:songjiajie2010